ORIGINAL ARTICLE
Year : 2017  |  Volume : 6  |  Issue : 3  |  Page : 296-301

Rapid laboratory diagnosis of pulmonary tuberculosis


Department of Microbiology, Grant Government Medical College and Sir JJ Group of Hospital, Mumbai, Maharashtra, India

Date of Web Publication31-Jul-2017

Correspondence Address:
Prasanna Bhirud
Department of Microbiology, Grant Government Medical College and Sir JJ Group of Hospital, Byculla, Mumbai - 400 008, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijmy.ijmy_61_17

Rights and Permissions
  Abstract 

Background: Tuberculosis (TB) ranks as the second leading cause of death from an infectious disease worldwide. Early diagnosis of Mycobacterium tuberculosis in clinical samples becomes important in the control of TB both for the treatment of patients and for curbing of disease transmission to the others in the community. The study objective was to perform Ziehl–Neelsen (ZN) staining, fluorochrome staining, line probe assay (LPA), and loop-mediated isothermal amplification (LAMP) assay for rapid detection of pulmonary TB (PTB) and to compare the results of LPA and LAMP in terms of sensitivity, specificity, and turnaround time. Methods: A total of 891 sputum samples from clinically diagnosed/suspected cases of TB were subjected to ZN and fluorochrome staining. Smear positive samples were subjected to LPA, and smear negative were cultured on Lowenstein–Jensen media. A total of 177 samples were subjected to liquid culture and LAMP. Conventional culture was considered as “gold standard” for calculation of parameters. Results: Light-emitting diode fluorescence microscopy had the same sensitivity as ZN with similar high specificity. LPA was performed on 548 sputum samples which includes 520 smear positive and 28 smear negative culture positive samples and multidrug-resistant TB was detected in 32.64%. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of TB-LAMP on direct sputum samples was found to be 98.96%, 95%, 96%, and 98.70%, respectively, when compared with ZN smear microscopy. By considering culture as “gold standard,” LAMP showed a sensitivity, specificity, PPV, and NPV of 98.94%, 96.34%, 96.90%, and 98.75%, respectively. The sensitivity and PPV of TB-LAMP were 98.97% and 96%, respectively, when compared with LPA. Conclusions: A successful rapid laboratory diagnosis of PTB is possible when one combines the available methodology of microscopy, culture as well as molecular techniques. The LAMP assay was found to be simple, self-contained, and efficacious for early diagnosis of suspected cases of PTB with advantages of having a high throughput, no requirements of sophisticated equipment, and complex biosafety facilities.

Keywords: Line probe assay, loop-mediated isothermal amplification, multidrug-resistant tuberculosis, tuberculosis


How to cite this article:
Bhirud P, Joshi A, Hirani N, Chowdhary A. Rapid laboratory diagnosis of pulmonary tuberculosis. Int J Mycobacteriol 2017;6:296-301

How to cite this URL:
Bhirud P, Joshi A, Hirani N, Chowdhary A. Rapid laboratory diagnosis of pulmonary tuberculosis. Int J Mycobacteriol [serial online] 2017 [cited 2023 Apr 1];6:296-301. Available from: https://www.ijmyco.org/text.asp?2017/6/3/296/211927


  Introduction Top


Tuberculosis (TB) ranks as the second leading cause of death from an infectious disease worldwide, after the human immunodeficiency virus. Worldwide, 9.6 million people are estimated to have fallen ill with TB in 2014. Globally, an estimated 3.3% of new TB cases and 20% of previously treated cases have been detected with multidrug-resistant (MDR) TB.[1] Early diagnosis of Mycobacterium tuberculosis (MTB) in clinical samples becomes important in the control of TB, both, for the treatment of patients and for curbing the disease transmission to others in the community.

Smear examination is believed to be simple, cheap, quick, practical, and an effective case finding method for developing countries. Microscopic examination of Ziehl–Neelsen (ZN) or auramine-stained specimen allows detection of most strains in less than an hour. The fluorochrome stain offers the advantage of greater sensitivity compared with carbolfuchsin method since a significantly larger area of the smear can be scanned per unit time with auramine fluorochrome stain.[2] In addition, a positive fluorescent smear may be restained using the conventional ZN or Kinyoun procedure.

Culture is considered the most accurate test due to its high sensitivity and specificity. However, this technique is both labor intensive and time consuming, requiring 6–8 weeks to achieve maximum sensitivity. In the last few years, various commercial molecular methods of nucleic acid amplification have been introduced for rapid TB diagnosis such as line probe assay (LPA), GeneXpert, and loop-mediated isothermal amplification (LAMP).

Recent advances in technology have introduced many rapid and reliable methods to differentiate between susceptible and resistant MTB strains; however, due to their high cost and equipment requirement, these new methods are not feasible in the clinical laboratories of developing countries in the diagnosis of TB. Instead, these countries use the proportional method, which is very time consuming. Consequently, physicians base their diagnosis of TB on microscopy results. Therefore, supplementary rapid and reliable methods are highly needed for clinical laboratories with limited resources. In 1995, Yajko et al.[3] used an oxidation-reduction indicator, Alamar Blue, which changes color in response to the growing bacteria. In 2004, Farnia et al. demonstrated the viability of MTB in sputum specimens of TB patients using malachite green indicator dye, a compound routinely used in Lowenstein–Jensen (LJ) medium. Malachite green is a triphenylmethane dye and has a dark green color, which becomes colorless during MTB metabolism.[4]

Kohli et al.'s study was aimed at comparing the indirect nitrate reductase assay (NRA) with the indirect proportion method in terms of speed, cost, ease of performance, and accuracy for drug susceptibility testing (DST) to first-line antitubercular drugs.[5]

Kammoun et al.'s study is to evaluate the performance of the NRA applied directly on microscopy-positive sputum samples from patients with pulmonary TB (PTB) for the detection of resistance to the first-line anti-TB drugs: rifampin (RIF), isoniazid (INH), streptomycin, and ethambutol.[6]

The LPA and GeneXpert also help in detection of drug resistance. The LPA is a DNA strip technology. The INNO-LIPA RIF-TB (Innogenetics, Ghent, Belgium) and and GenoType MTBDR (Hain Life-Science, Nehren, Germany) are available commercially. The INNO-LIPA RIF-TB detects rpoB gene and GenoType MTBDR detects rpoB, katG and inhA gene. Both tests are based on reverse hybridization of amplicons to immobilized membrane bound probes, allowing the detection of mutations at the level of the most frequently mutated codons. The presence of a mutation is revealed by the absence of hybridization at the level of the wild-type probes (rpoB WT1 to WT8 and katG-WT and inhAWT1 and 2), with a possible positive hybridization signal at the level of the mutant probes.

Recently, LAMP assay has been developed as a novel technique for nucleic acid amplification. This test is used for the detection of MTB complex which includes the following species: Mycobacterium africanum, Mycobacterium bovis, Mycobacterium canettii, Mycobacterium microti, and MTB.

LAMP can amplify DNA with high specificity and efficiency under isothermal conditions using six sets of primers that recognize eight distinct regions on the target sequence. Unlike polymerase chain reaction (PCR), LAMP reaction does not require a denatured DNA template and relies on autocycling strand displacement DNA synthesis by Bst DNA polymerase. The improved sensitivity is due to a large amount of amplicon generated by Bst polymerase. The large amount of DNA is generated in less than an hour, and a positive LAMP reaction can be visualized with the naked eye using a 254 nm wavelength ultraviolet (UV) transilluminator without the need for gel electrophoresis or any other instrumentations.[7]

Jeanes and O'Grady review outlined how next generation sequencing, through fast and accurate whole genome sequencing and/or shotgun metagenomics sequencing, holds the potential to transform the diagnosis and management of TB both for the individual patient and as an epidemiological tool.[8]


  Methods Top


This study was carried out at the Biosafety level III TB laboratory of a tertiary care hospital.

Study population and sample collection

A total of 891 patients were enrolled in this study over a period of 18 months. Clinically suspected/diagnosed PTB cases were included in the study while cases of extra PTB were excluded from the study. All sputum samples were received in the TB laboratory from clinically suspected/diagnosed cases of PTB. Two samples per patient as per RNTCP guidelines [9] were taken. These were subjected to direct smear by ZN and auramine staining.

Ziehl–Neelsen staining

Direct smears were made from the thick purulent part of sputum specimens, air dried, heat fixed, and stained by ZN staining method. The smears were scanned with an oil immersion objective and graded according to RNTCP guideline.[9]

Fluorochrome staining

Fluorescence stained smears were scanned under lower magnification and graded according to the International Union Against Tuberculosis Lung Disease grading using light-emitting diode (LED) fluorescent microscope (FM).

Sample processing

The sputum samples were decontaminated and concentrated using NALC–NaOH concentration method. Smear-positive samples were subjected to LPA while smear-negative samples were put up for culture using solid LJ media, and the culture-positive isolates were then subjected to LPA. A total of 177 samples (97 smear positive and 80 smear negative) were subjected to LAMP and liquid culture (mycobacterial growth indicator tube [MGIT]).

Liquid culture: Mycobacterial growth indicator tube

The MGIT works on the following principle: The MGIT tube has fluorescent compound embedded in silico ne at the bottom of the tube, and this fluorescent compound is sensitive to the presence of oxygen dissolved in the broth. Initially, the large amount of dissolved oxygen quenches emissions from the compound and little fluorescence can be detected. Later, actively respiring microorganisms consume the oxygen and allow the fluorescence to be detected. Methodology for MGIT was as per kit literature of BBL™ MGIT™.

Line probe assay

LPA is a DNA strip technology using a cellulose acetate membrane strip for the identification of MTB complex and detection of its resistance to RIF and/or INH. The methodology followed for LPA was as per kit literature of GenoType MTBDR plus version 2 kit (Hain Lifescience Germany). It is a qualitative in vitro test for the identification of the MTB complex and its resistance to rifampicin (RMP) and INH from pulmonary smear-positive clinical specimens and culture samples. The LPA involves DNA extraction from clinical specimens or cultured material, multiplex amplification with biotinylated primers and reverse hybridization. Interpretation of results is done using the reference template provided with the kit.[10]

Loop-mediated isothermal amplification

Nu-LAMP TB kit is an in vitro diagnostic LAMP test for detection of MTB complex. The assay also uses a set of four primers which recognize six regions on pathogen target which makes the assay highly specific. The target region selected in the pathogen genome is rpoB region of MTB complex. The whole procedure consists of two steps – DNA extraction from clinical specimen or cultured material and the LAMP reaction. The methodology for LAMP was followed as per kit literature using the Nu-LAMP TB kit of RAS Lifesciences Private Limited.[7]

DNA extraction

RAS-DNA extraction kit was used for the extraction. A 500 μl of the smear-positive decontaminated sputum specimen was used for the DNA extraction. The DNA extraction kit is a modifi cation of a procedure based on separating contaminating protein from DNA by salt precipitation. It involves digestion of cellular proteins, subsequent removal of proteins by salting out precipitation of DNA with isopropanol, and resuspension of buffer.[11]

Loop-mediated isothermal amplification reaction

The Working solution for LAMP reaction was prepared by using RAS-SYBR (fluorescent dye) and RAS-TEB (Dilution buffer for Fluorescent Dye). Separate PCR tubes used for each sample, positive control and negative control (MBGW). The master mix was prepared in a DNA-free room, and the DNA extract was added to the master mix in a PCR hood. Both these procedures were done in separate rooms following the restricted access and unidirectional workflow protocol. PCR tubes were placed in thermocycler and tubes were incubated at 65°C for 60 min. One microliter of working SYBR solution was added into each tube. The results were analyzed by visually observing the color change in the tube against black background and/or by observing the color change in the tube when kept against UV light at short range wavelength (254 nm) using fluorescence visualizer.

Statistical analysis

The results were analyzed using Open Epi software. P value was calculated by using Fisher exact test at the <0.05 level of significance. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated using culture as “gold standard.”


  Results Top


A total of 891 sputum specimens were received from patients attending the outpatient as well as inpatient departments of a tertiary care hospital from January 2014 to June 2015. Of these, 520 patients' samples were smear positive. Out of 520 patients, 375 were male and 145 were female. Age group 21–30 years showed maximum smear positivity (29.42%) followed by age group of 31–40 (25.20%).

Out of the total 891 samples, 520 were smear positive and 371 were smear negative by ZN and LED-based fluorescent microscopy, respectively. Out of the 520 smear positive samples, 59, 179, 135, and 147 samples were scanty, 1+, 2+, and 3+, respectively, by ZN staining method, while out of total 891 samples, 371 were smear negative. Fifty-six samples were graded as scanty; 156, 118, and 190 samples were graded as 1+, 2+, and 3+, respectively, by fluorochrome staining.

All 520 smear-positive sputum samples were put up for LPA; results of which are depicted in [Table 1].
Table 1: Line probe assay results of Ziehl–Neelsen smear-positive samples (n=520)

Click here to view


The 371 smear-negative samples were cultured on LJ media. Of these, 28 were culture positive. These smear-negative culture-positive isolates were subjected to LPA. Of the total 28 strains, 11 were detected as MDR, while 13 were sensitive to both INH and RMP. Three strains were resistant to INH.

Sputum samples from 177 clinically suspected cases of PTB were subjected to LAMP reaction. Of these, 97 were smear-positive and eighty were smear-negative samples.

Comparison of smear microscopy (ZN and fluorochrome staining), liquid culture, LPA, and LAMP is depicted in [Table 2].
Table 2: Comparison of Ziehl–Neelsen, culture, line probe assay, and loop-mediated isothermal amplification (n=177)

Click here to view


Out of the total eighty smear-negative samples, three smear negatives were positive by LAMP assay and were subjected to LPA (though smear-negative samples are not endorsed by World Health Organization [WHO] for LPA). MTB complex was detected in two of these, and they were also culture positive.

The sensitivity and PPV of LAMP assay were 98.97% and 96%, respectively, when compared with LPA. Specificity and NPV of LAMP assay were not calculated as LPA was not performed on smear-negative samples as per the WHO guidelines [Table 3].
Table 3: Comparison of smear microscopy (Ziehl–Neelsen and fluorochrome staining), liquid culture, line probe assay, and loop-mediated isothermal amplification

Click here to view



  Discussion Top


In our study, 520 out of the total 891 samples were smear positive; total smear-positive samples (58.36%) and majority of the patients (29.42%) were in the age group of 21–30 years, followed by 25.20% being in the age group of 31–40 years with 72.12% being male and 27.88% female patients. According to RNTCP status report 2012,[9] TB primarily affects people in their most reproductive years of life. Our study is in concordance with the same.

The M.tb fluoresces as greenish under fluorescent microscope as shown in [Figure 1].
Figure 1: Fluorescent microscopy of positive sputum sample

Click here to view


In 2010, the WHO recommended conventional FM to be replaced by LED-FM and that LED microscopy be phased in as an alternative for conventional ZN microscopy.[12] In this study, LED-FM had the same sensitivity as compared with ZN and similar specificity. A study by Bonnet et al. also showed that LED-FM had the same sensitivity as ZN with a similar high specificity.[13] In our study, the specificity was not affected by the use of a very sensitive acid-fast bacilli (AFB) cutoff (1–10 AFB/field) to define a positive smear, and the proportion of scanty results was 10.77% among positive smears. Smear reading was faster with LED-FM compared to conventional ZN microscopy. This is consistent with the 25%–66% time saving when using FM compared to ZN microscopy as reported by Marais et al. study.[14] Regarding feasibility aspects, this study confirmed the very good acceptability of the LED-FM, the long working life of the diodes in the conditions of a peripheral laboratory, and the possibility to use LED-FM without a dark room.

In Bhalla et al. study, although differences were not statistically significant, highest sensitivity was achieved with LED-FM with a good overall agreement of 82.6%. A total of 200 patients were included in this study. Sensitivity and specificity for the LED assessment, MVP assessment, and light microscopy were 83.1% and 82.4%, 78.5% and 87.5%, and 81.6% and 83.5%, respectively. Mean reading time was approximately 3 times faster than ZN microscopy. The mean time to read a negative smear was 2 min with FM and 5 min with light microscopy with time savings of 60%.[15]

In our study, LPA was performed on a total of 548 samples which included smear-positive samples and culture-positive isolates. MTB complex was not detected in 18 samples. MDR-TB was detected in 32.64% which was very high.

The LPA is directly performed on clinical samples and offers an enormous advantage to the patients. The turnaround time of LPA (24–48 h) is very short as compared with conventional DST methods, and hence, it would be highly advantageous for high-burden areas.

Various nucleic acid amplification-based methods have been developed to address the need for rapid and sensitive diagnosis of MTB and other mycobacterial infections. These methods require either sophisticated instruments for the amplification or elaborate methods for detection of the amplified products, which are major obstacles.

Application of LAMP to clinical specimens was evaluated by comparing LAMP results with acid-fast smear tests, liquid culture, and LPA. The LAMP showed excellent sensitivity 98.97%, specificity 95%, PPV 96%, and NPV 98.70% when compared with liquid culture by considering it as “gold standard.”

The LAMP reaction results were analysed by observing the color change in the tube when kept against UV light at short range wavelength (254 nm) using fluorescence visualize as shown in [Figure 2].
Figure 2: LAMP assay showing greenish fluorescence of Mycobacterium tuberculosis positive sputum samples

Click here to view


The sensitivity of LAMP in smear-positive and culture-positive samples was 98.94% (94/95). Smear-negative culture-positive sample size was very low (n = 2) and both samples were detected by LAMP. Another multicenter evaluation study involving 1061 patients tested in reference laboratories in Vietnam, South Africa, Peru, and Brazil showed that TB-LAMP detected almost 97% of smear-positive/culture-positive patients and 53% of smear-negative/culture-positive patients.[16]

The detection rate of LPA was found to be 54.80% (97/177) whereas LAMP was found to be 56.50% (100/177). It was 1.7% higher in case of LAMP assay. In a study by Kumar et al., LAMP assay was able to detect 8.5% additional cases over mPCR.[17] Mitarai et al. reported that the sensitivity of direct LAMP (88.2%) was slightly but not statistically significantly lower than Cobas Amplicor MTB and TRC Rapid MTB.[18]

In Safi et al. study, sensitivity of LAMP assay for detection of DNA of Mycobacterium avium subspecies paratuberculosis (MAP) was 4 fg/ll, and the specificity was 100%. This assay successfully detected MAP not only in the bacterial cultures but also in clinical fecal samples and the specificity of both PCR was 100%.[19]

In Sanker et al. study, Xpert and LPA have a mismatch of up to 20% with MGIT (RIF 0.5 mg/L); the concerned mutations are more frequently encountered toward both ends of rpoB RDR among the tested samples. The increased frequency of mutations associated with “sub-breakpoint low-level resistance,” hetero-resistant bacterial populations, and inherent limitations of the nucleic acid amplification tests (NAATs) and MGIT are probably responsible for the mismatch among samples. Inadequate drug exposure due to low-RIF dosages, poor directly observed therapy, and other issues associated with poor serum drug levels along with the low relative risk (RR) prevalence are the probable reasons. Extensive use of NAATs among such a population without a strictly targeted approach and confirmatory phenotypic tests with multiple RIF concentrations may result in false RR results on a massive scale.[20]

These results indicate that LAMP is a highly sensitive assay and can detect small quantity of mycobacterial DNA in clinical samples.

LAMP was compared with LPA in terms of turnaround time for detection of MTB complex, wherein LPA takes 24–48 h, whereas LAMP results are obtained within 3 h. The advantages of LAMP assay over LPA are that the LAMP assay does not require the use of sophisticated instruments and helps in rapid detection of MTB.

The only advantage of LPA over LAMP assay is that it gives DST results, whereas the LAMP assay can only detect MTB complex.


  Conclusions Top


A successful rapid laboratory diagnosis of PTB is possible when one combines available methodology of microscopy, culture as well as molecular techniques.

LED-based fluorescent microscopy did not have an increased sensitivity or specificity as compared to ZN microscopy, but the simplicity of staining method combined with faster viewing time, long shelf life of the LED bulbs, and the battery backup gives the method an advantage over ZN microscopy.

LPA performed on smear-positive sputum samples was sensitive and specific for detection of PTB, with the added advantage of drug sensitivity testing for RMP and INH. Furthermore, the rapid turnaround time of 48 h helps in immediate initiation of treatment.

LAMP is a rapid molecular test for detection of MTB from pulmonary samples and useful when resources limited, prevalence of TB infection high but prevalence of MDR-TB low. In Mumbai, utility of LAMP is limited as the prevalence of MDR-TB is high (35%–45%) though LAMP has a rapid turnaround time and high sensitivity; LPA was found to be highly sensitive and specific with an advantage of DST within 48 h.

Acknowledgments

The authors would like to thank Central TB division and state TB office for funding, technical staff of TB culture and DST laboratory for providing technical support.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
World Health Organization. Global Tuberculosis Report 2015. Geneva: World Health Organization; 2015. Available from: http://www.who.int/ tb/publications/global_report/gtbr15_main_text.pdf. [Last accessed on 2017 May 25].  Back to cited text no. 1
    
2.
Winn WC, Koneman EW, Allen SD, Procop GW, Janda WM, Woods GL, editors. Mycobacteria In: Koneman's Colour Atlas and Textbook of Diagnostic Microbiology. 6th ed. Philadelphia: Lippincott Williams and Wilkins; 2006. p. 1065-124.  Back to cited text no. 2
    
3.
Yajko DM, Madej JJ, Lancaster MV, Sanders CA, Cawthon VL, Gee B, et al. Colorimetric method for determining MICs of antimicrobial agents for Mycobacterium tuberculosis. J Clin Microbiol 1995;33:2324-7.  Back to cited text no. 3
    
4.
Farnia P, Masjedi MR, Mohammadi F, Tabarsei P, Farnia P, Mohammadzadeh AR, et al. Colorimetric detection of multidrug-resistant or extensively drug-resistant tuberculosis by use of malachite green indicator dye. J Clin Microbiol 2008;46:796-9.  Back to cited text no. 4
    
5.
Kohli A, Bashir G, Fatima A, Jan A, Wani NU, Ahmad J. Rapid drug-susceptibility testing of Mycobacterium tuberculosis clinical isolates to first-line antitubercular drugs by nitrate reductase assay: A comparison with proportion method. Int J Mycobacteriol 2016;5:469-74.  Back to cited text no. 5
  [Full text]  
6.
Kammoun S, Smaoui S, Marouane C, Slim L, Messadi-Akrout F. Drug susceptibility testing of Mycobacterium tuberculosis by a nitrate reductase assay applied directly on microscopy-positive sputum samples. Int J Mycobacteriol 2015;4:202-6.  Back to cited text no. 6
  [Full text]  
7.
Nu LAMP TB kit (product insert) (Loop mediated isothermal amplification TB Test Kit based on Real Time PCR/Visual Detection) RAS Lifesciences Pvt. Ltd. Available from www.raslifesciences.com. [Last accessed on 2017 May 25].  Back to cited text no. 7
    
8.
Jeanes C, O'Grady J. Diagnosing tuberculosis in the 21st century-Dawn of a genomics revolution? Int J Mycobacteriol 2016;5:384-91.  Back to cited text no. 8
  [Full text]  
9.
Revised National Tuberculosis Control Programme: RNTCP at A Glance. Published By Central TB Division, Ministry of Health and Family Welfare; 2012.  Back to cited text no. 9
    
10.
GenoType MTBDRplus, Version 2.0. (Product Insert). Nehren, Germany: Hain Lifescience GmbH; February, 2012. Available from: http://www.hain-lifescience.com.pdf. [Last accessed on 2017 May 25].  Back to cited text no. 10
    
11.
RAS DNA Extraction Kit (product insert) RAS Lifesciences Pvt. Ltd. Available from: www.raslifesciences.com. [Last accessed on 2017 May 25].  Back to cited text no. 11
    
12.
World Health Organization. Available from: http://www.who.int/tb/advisory_bodies/stag_tb_report_2009.pdf. [Last accessed on 2011 Jan 26].  Back to cited text no. 12
    
13.
Bonnet M, Gagnidze L, Githui W, Guérin PJ, Bonte L, Varaine F, et al. Performance of LED-based fluorescence microscopy to diagnose tuberculosis in a peripheral health centre in Nairobi. PLoS One 2011;6:e17214.  Back to cited text no. 13
    
14.
Marais BJ, Brittle W, Painczyk K, Hesseling AC, Beyers N, Wasserman E, et al. Use of light-emitting diode fluorescence microscopy to detect acid-fast bacilli in sputum. Clin Infect Dis 2008;47:203-7.  Back to cited text no. 14
    
15.
Bhalla M, Sidiq Z, Sharma PP, Singhal R, Myneedu VP, Sarin R. Performance of light-emitting diode fluorescence microscope for diagnosis of tuberculosis. Int J Mycobacteriol 2013;2:174-8.  Back to cited text no. 15
  [Full text]  
16.
WHO: The Use of a Commercial Loop-mediated Isothermal Amplification Assay (TB-LAMP) for the Detection of Tuberculosis Expert Group Meeting Report Geneva; May, 2013.  Back to cited text no. 16
    
17.
Kumar P, Pandya D, Singh N, Behera D, Aggarwal P, Singh S. Loop-mediated isothermal amplification assay for rapid and sensitive diagnosis of tuberculosis. J Infect 2014;69:607-15.  Back to cited text no. 17
    
18.
Mitarai S, Okumura M, Toyota E, Yoshiyama T, Aono A, Sejimo A, et al. Evaluation of a simple loop-mediated isothermal amplification test kit for the diagnosis of tuberculosis. Int J Tuberc Lung Dis 2011;15:1211-7, i.  Back to cited text no. 18
    
19.
Safi S, Heidarnejhad O, Mosavari N, Sakha M, Afshar D, Moazami L, et al. Comparative evaluation of LAMP and Nested-PCR for the diagnosis of bovine paratuberculosis. Int J Mycobacteriol 2015;4 Suppl 1:98-9.  Back to cited text no. 19
    
20.
Sanker P, Ambika AP, Santhosh VT, Kottuthodi RP, Balakrishnan R, Mrithunjayan SK, et al. Are WHO approved nucleic acid amplification tests causing large-scale “false identification” of rifampicin-resistant tuberculosis? Programmatic experience from South India. Int J Mycobacteriol 2017;6:21-6.  Back to cited text no. 20
[PUBMED]  [Full text]  


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Value analysis of next-generation sequencing combined with Xpert in early precise diagnosis of pulmonary tuberculosis
Xingwu Zou, Yanling Zhu, Yao Qin, Fan Fei, Yuanyuan Chen, Pengcheng Wang, Lihong Zhou, Yazhen Lang
Diagnostic Microbiology and Infectious Disease. 2023; : 115921
[Pubmed] | [DOI]
2 Analysis of the application value of molecular diagnostic technology of Mycobacterium tuberculosis
Yinyin Zhu, Rong Wang, Mengkai Qiao, Minghe Guo, Yan Kong, Hongying Zhang
Journal of Clinical Tuberculosis and Other Mycobacterial Diseases. 2022; : 100330
[Pubmed] | [DOI]
3 Evolution of Machine Learning in Tuberculosis Diagnosis: A Review of Deep Learning-Based Medical Applications
Manisha Singh, Gurubasavaraj Veeranna Pujar, Sethu Arun Kumar, Meduri Bhagyalalitha, Handattu Shankaranarayana Akshatha, Belal Abuhaija, Anas Ratib Alsoud, Laith Abualigah, Narasimha M. Beeraka, Amir H. Gandomi
Electronics. 2022; 11(17): 2634
[Pubmed] | [DOI]
4 Diagnostic utility of GenoType MTBDRsl assay for the detection of moxifloxacin-resistant mycobacterium tuberculosis, as compared to phenotypic method and whole-genome sequencing
RajNarayan Yadav, Manpreet Bhalla, Gavish Kumar, GrishC Sah, RavindraKumar Dewan, Ritu Singhal
International Journal of Mycobacteriology. 2022; 11(2): 183
[Pubmed] | [DOI]
5 Diagnostic challenges and Gene-Xpert utility in detecting Mycobacterium tuberculosis among suspected cases of Pulmonary tuberculosis
Shaila Kabir,M. Tanveer Hossain Parash,Nor Amalina Emran,A. B. M. Tofazzal Hossain,Sadia Choudhury Shimmi,Frederick Quinn
PLOS ONE. 2021; 16(5): e0251858
[Pubmed] | [DOI]
6 Rapid molecular diagnostics for multi-drug resistant tuberculosis in India
Rajeswari Ramachandran,M Muniyandi
Expert Review of Anti-infective Therapy. 2018; : 1
[Pubmed] | [DOI]
7 Diagnostic performance of GeneXpert and Ziehl-Neelson microscopy in the detection of tuberculosis in Benue State, Nigeria
Enenche Francis Ejeh,Ann Undiandeye,Victor O. Akinseye,Kenneth O. Okon,Haruna M. Kazeem,Caleb Ayuba Kudi,Simeon Idowu B. Cadmus
Alexandria Journal of Medicine. 2018;
[Pubmed] | [DOI]
8 Parallel Tests Using Culture, Xpert MTB/RIF, and SAT-TB in Sputum Plus Bronchial Alveolar Lavage Fluid Significantly Increase Diagnostic Performance of Smear-Negative Pulmonary Tuberculosis
Lin Fan,Danfeng Li,Shaojun Zhang,Lan Yao,Xiaohui Hao,Jin Gu,Hong Li,Jinxia Niu,Zhemin Zhang,Changtai Zhu
Frontiers in Microbiology. 2018; 9
[Pubmed] | [DOI]
9 Evaluation of the RT-LAMP and LAMP methods for detection of Mycobacterium tuberculosis
Dandan Wu,Jiwen Kang,Baosheng Li,Dianxing Sun
Journal of Clinical Laboratory Analysis. 2017; : e22326
[Pubmed] | [DOI]
10 Identification and comparison of novel circular RNAs with associated co-expression and competing endogenous RNA networks in pulmonary tuberculosis
Xing Zhang,Min Zhu,Rong Yang,Weifeng Zhao,Xiaolong Hu,Jianhe Gan
Oncotarget. 2017; 8(69): 113571
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methods
Results
Discussion
Conclusions
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed7254    
    Printed214    
    Emailed1    
    PDF Downloaded1063    
    Comments [Add]    
    Cited by others 10    

Recommend this journal